Endress+Hauser SS3000
Endress+Hauser SS3000 - Optical analysis - Degree of protection: Type 3R; Channels: 2; Measuring principle: TDLAS.
Endress+Hauser SS3000 is a Optical analysis catalog item from Endress+Hauser. The manufacturer's published data for this part number specifies Degree of protection: Type 3R and Channels: 2.
Analog Output: 1 or 2 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C Protocol: Modbus Gould RTU or Daniel RTU or ASCII Alarms: Four General Fault and Concentration Alarms via Modbus and Analog Output(s)
The specification set recorded for this part number covers Measuring principle: TDLAS; Hazardous area approvals: CSA Class I, Division 2 CSA Class I, Zone 2; Measured variables: Concentration Moisture Dew Point Cell Pressure Cell Temperature; Ambient temperature range: -20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F); Operating Pressure Range: Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 700 to 1400 mbara; Analyte and measurement ranges: H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum); Analyzer wetted materials: 316L stainless steel FKM O-Rings Glass; Housing materials: Electronics: 304 stainless steel Sample System Panel: Anodized aluminum. The complete technical table is shown in the Specifications tab on this page.
| Ambient temperature range | -20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F) |
|---|---|
| Analyte and measurement ranges | H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum) |
| Analyzer wetted materials | 316L stainless steel FKM O-Rings Glass |
| Channels | 2 |
| Communication | Analog Output: 1 or 2 4-20mA Isolated, 1200 ohms at 24 VDC max load Serial: RS232C Protocol: Modbus Gould RTU or Daniel RTU or ASCII Alarms: Four General Fault and Concentration Alarms via Modbus and Analog Output(s) |
| Degree of protection | Type 3R |
| Hazardous area approvals | CSA Class I, Division 2 CSA Class I, Zone 2 |
| Housing materials | Electronics: 304 stainless steel Sample System Panel: Anodized aluminum |
| Measured variables | Concentration Moisture Dew Point Cell Pressure Cell Temperature |
| Measuring principle | TDLAS |
| Operating Pressure Range | Inlet Pressure: 140 to 350 kPa (20 to 50 psig) Sample Cell: 700 to 1400 mbara |
| Power Supply | 100 to 240 VAC, 50 to 60 Hz or 9 to 16 VDC or 18 to 32 VDC - optional 1 Amp maximum at 120 VAC 1.6 Amps at 24 VDC, 3.2 Amps at 12 VDC |
| Product headline | Dual channel gas analyzer for two moisture (H2O) channels. The sensor measures gas using tunable diode laser absorption spectroscopy (TDLAS) to determine the concentration of the gas without coming into physical contact with the stream. |
Frequently Asked Questions
How do I get pricing for SS3000?
KKM Solutions is a quote-based industrial supplier, so prices are not listed online. Send the quote form on this page with your quantity, or email orderdesk@kkmsolutions.com, and our order desk will reply with pricing and availability within 1 business day.
Who manufactures the SS3000?
SS3000 is a Endress+Hauser part. KKM Solutions sources Endress+Hauser components to the exact manufacturer specification and cross-checks every order against manufacturer part data before it ships.
What are the key specifications of SS3000?
Key specifications include Ambient temperature range: -20 to 50°C (-4 to 122°F) Optional: -10 to 60°C (14 to 140°F); Analyte and measurement ranges: H2O (Moisture): 0 to 100 ppmv (minimum) 0 to 5000 ppmv (maximum); Analyzer wetted materials: 316L stainless steel FKM O-Rings Glass; Channels: 2. The full specification table on this page lists every published attribute for this part.
How quickly can SS3000 ship?
In-stock items are dispatched the same business day for orders placed by 3 PM; sourced and quoted items ship per the quoted lead time. KKM Solutions ships worldwide — to your site, warehouse, or freight forwarder.
Related Products
Endress+Hauser Endress+Hauser SS2100i gas analyzer
Endress+Hauser Endress+Hauser SS3000E
Endress+Hauser Endress+Hauser SS2100A
Endress+Hauser Endress+Hauser SS500 moisture analyzer
Endress+Hauser Endress+Hauser SS500E
Endress+Hauser
Industrial & MRO Parts