Endress+Hauser HydrostaticLevel measurementDeltapilot FMB51
Endress+Hauser FMB51 - Pressure - Ambient temperature: -40°C...85°C (-40°F...185°F); Material housing: 316L, Die-cast aluminum; Options: Separate housing Initial device settings Overvoltage protection.
The Endress+Hauser FMB51 is part of the Pressure range in our industrial catalog. The manufacturer's published data for this part number specifies Ambient temperature: -40°C...85°C (-40°F...185°F) and Material housing: 316L, Die-cast aluminum.
4...20 mA HART 10,5...45 VDC (Non Ex): Ex ia: 10,5...30 VDC PROFIBUS PA: 9...32 VDC (Non Ex) FOUNDATION Fieldbus: 9...32 VDC (Non Ex)
Key manufacturer data on file for this item includes Options: Separate housing Initial device settings Overvoltage protection; Communication: 4...20 mA HART PROFIBUS PA FOUNDATION Fieldbus; Accuracy: Standard 0.2% Optional 0.1%; Measuring principle: Hydrostatic; Process Connection: Thread: G1 1/2, MNPT1 1/2 Flange: DN40...DN100, ASME 2"...4", JIS 10K; Process temperature: -10°C...85°C (14°F…212°F); Process pressure / max. overpressure limit: 40 bar (600 psi); Main wetted parts: Alloy C 316L optional coating AuPt; Certificates / Approvals: ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI; Design approvals: EN 10204-3.1 NACE MR0175; Max. measurement distance: 100m (328ft) H2O; Reference Accuracy: Standard 0.2 % Optional 0.1 %. The complete technical table is shown in the Specifications tab on this page.
| Accuracy | Standard 0.2% Optional 0.1% |
|---|---|
| Ambient temperature | -40°C...85°C (-40°F...185°F) |
| Application limits | If pressurized, usage of two pressure transmitters to measure the differential pressure (electronic dP) Observe ratio head pressure : hydrostatic pressure |
| Certificates / Approvals | ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI |
| Characteristic | Pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell: Hermetically sealed, condensate-resistant, climatic-proofed and with lowest temperature influences Rod version |
| Characteristic / Application | Pressure transmitter for hydrostatic level measurement with flush mounted metallic Contite measuring cell: Hermetically sealed, condensate-resistant and climatic-proofed Rod version |
| Communication | 4...20 mA HART PROFIBUS PA FOUNDATION Fieldbus |
| Design approvals | EN 10204-3.1 NACE MR0175 |
| Drinking water approvals | KTW/ NSF |
| Fill fluid | Inert oil, Synthetic oil |
| Long term stability | 0.05 % of URL/ year |
| Main wetted parts | Alloy C 316L optional coating AuPt |
| Material gasket | Viton, EPDM, Kalrez, none |
| Material housing | 316L, Die-cast aluminum |
| Material process membrane | 316L, AlloyC Gold-Rhodium |
| Max. measurement distance | 100m (328ft) H2O |
| Max. overpressure limit | 40 bar (600 psi) |
| Max. Turn down | 100:1 |
| Measuring cell | 100 mbar...10 bar (1.5 psi...150 psi) |
| Measuring principle | Hydrostatic |
| Options | Separate housing Initial device settings Overvoltage protection |
| Pressure measuring range | 100 mbar...10 bar (1.5 psi...150 psi) |
| Process Connection | Thread: G1 1/2, MNPT1 1/2 Flange: DN40...DN100, ASME 2"...4", JIS 10K |
| Process pressure / max. overpressure limit | 40 bar (600 psi) |
| Process temperature | -10°C...85°C (14°F…212°F) |
| Reference Accuracy | Standard 0.2 % Optional 0.1 % |
| Safety approvals | Overfill protection WHG SIL |
| Smallest calibratable span | 10 mbar (1.45 psi) |
| Specialities | Modularity to differential pressure and process pressure devices (replacable display, universal electronics) diagnostic functions Hermetically sealed Contite measuring cell (condensate-resistant and climatic-proofed) |
| Supply / Communication | 4...20mA HART: 10,5...45V DC Ex ia: 10,5...30V DC PROFIBUS PA / FOUNDATION Fieldbus: 9...32V DC (Non Ex) |
| Supply voltage | 4...20 mA HART 10,5...45 VDC (Non Ex): Ex ia: 10,5...30 VDC PROFIBUS PA: 9...32 VDC (Non Ex) FOUNDATION Fieldbus: 9...32 VDC (Non Ex) |
| Vacuum resistance | 0 mbar abs. |
Frequently Asked Questions
How do I get pricing for FMB51?
KKM Solutions is a quote-based industrial supplier, so prices are not listed online. Send the quote form on this page with your quantity, or email orderdesk@kkmsolutions.com, and our order desk will reply with pricing and availability within 1 business day.
Who manufactures the FMB51?
FMB51 is a Endress+Hauser part. KKM Solutions sources Endress+Hauser components to the exact manufacturer specification and cross-checks every order against manufacturer part data before it ships.
What are the key specifications of FMB51?
Key specifications include Accuracy: Standard 0.2% Optional 0.1%; Ambient temperature: -40°C...85°C (-40°F...185°F); Application limits: If pressurized, usage of two pressure transmitters to measure the differential pressure (electronic dP) Observe ratio head pressure : hydrostatic pressure; Certificates / Approvals: ATEX, FM, CSA, CSA C/US, IEC Ex, INMETRO, NEPSI. The full specification table on this page lists every published attribute for this part.
How quickly can FMB51 ship?
In-stock items are dispatched the same business day for orders placed by 3 PM; sourced and quoted items ship per the quoted lead time. KKM Solutions ships worldwide — to your site, warehouse, or freight forwarder.
Related Products
Endress+Hauser Endress+Hauser Hydrostatic Level measurementDeltapilot FMB50
Endress+Hauser Endress+Hauser HydrostaticLevel measurementDeltapilot FMB52
Endress+Hauser Endress+Hauser HydrostaticLevel measurementDeltapilot FMB53
Endress+Hauser Endress+Hauser HydrostaticLevel measurementDeltapilot FMB70
Endress+Hauser Endress+Hauser Electronicdifferential pressureDeltabar FMD71
Endress+Hauser
Industrial & MRO Parts